Rigol DM3058 Specifikace Strana 85

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 264
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 84
RIGOL
Programming Guide for DM3058
3-59
16. :calculate:limit:lower
Command Format:
:calculate:limit:lower?
:calculate:limit:lower<value>
Function:
The command queries and sets lower value of P/F calculation.
Parameters:
Name
Type
Range of Value
Default
<value>
consecutive real number
See Table 3-11
0
Explanations:
:calculate:limit:lower?: Query the lower value of P/F calculation.
:calculate:limit:lower<value>: Set the
lower value of P/F calculation. The
default range of <range> please refer to table 3-11.
Table 3-11 The range of P/F calculation
Measurement
Range
DEF
Unit
DC Voltage
± 1200
0
V
AC Voltage
0~900
0
V
DC Current
±12
0
A
AC Current
0~12
0
A
Resistance
0~1.2e+08
0
Ω
Capacitance
0~1.2e-02
0
F
Frequency
0~1.2e+06
0
HZ
Return Format:
The query returns current lower value.
17. :calculate:limit:upper
Command Format:
:calculate:limit:upper?
:calculate:limit:upper<value>
Zobrazit stránku 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 263 264