RIGOL
Programming Guide for DM3058
Explanations:
[SENSe:]RESistance:RANGe:AUTO □ {ON|OFF}: Enable or disable 2WR
autorange measuring function. If 2WR autorange function is ON, the measuring
range will be selected according to input signal.
[SENSe:]RESistance:RANGe:AUTO?: Query the enable state of 2WR autorange
measuring function.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The command returns the enable state of 2WR autorange measuring function.
40. [SENSe:]RESistance:RANGe[:UPPer]
Command Format:
[SENSe:]RESistance:RANGe[:UPPer]□{<range>|MIN|MAX|DEF}
[SENSe:]RESistance:RANGe[:UPPer]?□[{MIN|MAX}]
Function:
The command sets or queries the range of 2WR measurement.
Parameters:
{200 Ω | 2 kΩ | 20 kΩ | 200 kΩ
| 2 MΩ | 10 MΩ | 100 MΩ}
Explanations:
[SENSe:]RESistance:RANGe[:UPPer]□{<range>|MIN|MAX|DEF}: Set the range
of 2WR measurement. <range> may be 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 10
MΩ or 100 MΩ, MIN = 200 Ω, MAX =100 MΩ. Unit is Ω, so 2 kΩ should be input in
form of 2000.
[SENSe:]RESistance:RANGe[:UPPer]?□[{MIN|MAX}]: Query the range of 2WR
measurement.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
Komentáře k této Příručce