RIGOL
Programming Guide for DM3058
Function:
The command sets or queries the enable state of capacitance NULL measuring
function.
Explanations:
[SENSe:]CAPacitance:NULL[:STATe]□{ON|OFF}: Enable or disable capacitance
NULL measuring function.
[SENSe:]CAPacitance:NULL[:STATe]?: Query the enable state of capacitance
NULL measuring function.
The command ―[SENSe:]CAPacitance:NULL:VALue‖ can be used to set NULL
value.
The commands ―CONFigure:CAPacitance‖ and ―MEASure:CAPacitance?‖ can be
used to disable NULL function (OFF).
Return Format:
The query returns the enable state of capacitance measuring NULL function.
63. [SENSe:]CAPacitance:NULL:VALue
Command Format:
[SENSe:]CAPacitance:NULL:VALue□{<value>|MIN|MAX}
[SENSe:]CAPacitance:NULL:VALue?□[{MIN|MAX}]
Function:
The command sets or queries the NULL value of capacitance measurement.
Parameters:
Explanations:
[SENSe:]CAPacitance:NULL:VALue□{<value>|MIN|MAX}: Set the NULL value of
capacitance measurement. MIN is -120% max range of current function, MAX is
120% max range of current function, DEF = 0.
[SENSe:]CAPacitance:NULL:VALue?□[{MIN|MAX}]: Query the NULL value of
capacitance measurement.
Before sending the command, do run the command ―[SENSe:]CAPacitance:NULL
Komentáře k této Příručce