RIGOL
Programming Guide for DM3058
multiples of PLC.
Parameters:
{0.006|0.02|0.06|0.2|1|2|10|100}
Explanations:
[SENSe:]VOLTage[:DC]:NPLC□{<PLCs>|MIN|MAX|DEF}: Set the interval time of
DCV measurement, input value is multiples of PLC. <PLCs> may be 0.006, 0.02,
0.06, 0.2, 1, 2, 10 or 100, MIN = 0.006 PLC, MAX = 100 PLC.
[SENSe:]VOLTage[:DC]:NPLC?□[{MIN|MAX}]: Query the interval time of DCV
measurement, return value is multiples of PLC.
Return Format:
The query returns the interval time of DCV measurement which is multiples of
PLC.
26. [SENSe:]VOLTage[:DC]:NULL[:STATe]
Command Format:
[SENSe:]VOLTage[:DC]:NULL[:STATe]□{ON|OFF}
[SENSe:]VOLTage[:DC]:NULL[:STATe]?
Function:
The command sets or queries the enable state of DCV NULL function.
Explanations:
[SENSe:]VOLTage[:DC]:NULL[:STATe]□{ON|OFF}: Enable or disable DCV NULL
function.
[SENSe:]VOLTage[:DC]:NULL[:STATe]?: Query the enable state of DCV NULL
function.
The command ―[SENSe:]VOLTage[:DC]:NULL:VALue‖ can be used to set NULL
value.
The commands ―CONFigure:VOLTage:DC‖ and ―MEASure:VOLTage:DC?‖ can be
used to disable NULL function (OFF).
Return Format: