Rigol DM3058 Specifikace Strana 198

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 264
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 197
RIGOL
Programming Guide for DM3058
120%max range of current function, DEF = 0.
[SENSe:]FREQuency:NULL:VALue? [{MIN|MAX}]: Query the NULL value of
frequency measurement.
Before sending the command, do run the command [SENSe:]FREQuency:NULL
[:STATe]ON to enable NULL function.
Return Format:
The query returns the NULL value of frequency measurement.
55. [SENSe:]FREQuency:VOLTage:RANGe:AUTO
Command Format:
[SENSe:]FREQuency:VOLTage:RANGe:AUTO{ON|OFF}
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
Function:
The command sets or queries the enable state of voltage autorange measuring
function in frequency measurement.
Explanations:
[SENSe:]FREQuency:VOLTage:RANGe:AUTO {ON|OFF}: Enble or disable
voltage autorange measuring function in frequency measurement.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?: Query the enable state of voltage
autorange measuring function in frequency measurement.
The max range of AC voltage is 750 V (rms), voltage RMS is determined by
waveform.
Return Format:
The query returns the enable state of voltage autorange measuring function in
frequency measurement.
56. [SENSe:]FREQuency:VOLTage:RANGe[:UPPer]
Command Format:
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]{<range>|MIN|MAX|DEF}
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]?[{MIN|MAX}]
Zobrazit stránku 197
1 2 ... 193 194 195 196 197 198 199 200 201 202 203 ... 263 264