RIGOL
Programming Guide for DM3058
120%max range of current function, DEF = 0.
[SENSe:]FREQuency:NULL:VALue? □ [{MIN|MAX}]: Query the NULL value of
frequency measurement.
Before sending the command, do run the command ―[SENSe:]FREQuency:NULL
[:STATe]□ON‖ to enable NULL function.
Return Format:
The query returns the NULL value of frequency measurement.
55. [SENSe:]FREQuency:VOLTage:RANGe:AUTO
Command Format:
[SENSe:]FREQuency:VOLTage:RANGe:AUTO□{ON|OFF}
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
Function:
The command sets or queries the enable state of voltage autorange measuring
function in frequency measurement.
Explanations:
[SENSe:]FREQuency:VOLTage:RANGe:AUTO □ {ON|OFF}: Enble or disable
voltage autorange measuring function in frequency measurement.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?: Query the enable state of voltage
autorange measuring function in frequency measurement.
The max range of AC voltage is 750 V (rms), voltage RMS is determined by
waveform.
Return Format:
The query returns the enable state of voltage autorange measuring function in
frequency measurement.
56. [SENSe:]FREQuency:VOLTage:RANGe[:UPPer]
Command Format:
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]□{<range>|MIN|MAX|DEF}
[SENSe:]FREQuency:VOLTage:RANGe[:UPPer]?□[{MIN|MAX}]