Programming Guide for DM3058
The query returns the enable state of voltage autorange measuring function in
period measurement.
61. [SENSe:]PERiod:VOLTage:RANGe[:UPPer]
Command Format:
[SENSe:]PERiod:VOLTage:RANGe[:UPPer]{<range>|MIN|MAX|DEF}
[SENSe:]PERiod:VOLTage:RANGe[:UPPer]?[{MIN|MAX}]
Function:
The command sets or queries the voltage range of period measurement.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 mV | 2 V | 20 V
| 200 V | 750 V}
20 V
Explanations:
[SENSe:]PERiod:VOLTage:RANGe[:UPPer] {<range>|MIN|MAX|DEF}: Set the
voltage range of period measurement. <range> may be 200 mV, 2 V, 20 V, 200 V
or 750 V, MIN = 200 mV, MAX = 750 V. Unit is V, so 200 mV should be input in
form of 0.2.
[SENSe:]PERiod:VOLTage:RANGe[:UPPer]? [{MIN|MAX}]: Query the voltage
range of period measurement.
The max range of AC voltage is 750 V (rms), voltage RMS is determined by
waveform.
Return Format:
The query returns the voltage range of period measurement.
62. [SENSe:]CAPacitance:NULL[:STATe]
Command Format:
[SENSe:]CAPacitance:NULL[:STATe]{ON|OFF}
[SENSe:]CAPacitance:NULL[:STATe]?