RIGOL
Programming Guide for DM3058
Parameters:
Explanations:
[SENSe:]RESistance:APERture □ {<seconds>|MIN|MAX|DEF}: Set the interval
time of 2-wire resistance measurement. Unit is s, so ―400 ms‖ should be input in
form of 0.4.
[SENSe:]RESistance:APERture?□[{MIN|MAX}]: Query the interval time of 2-wire
resistance measurement.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the interval time of 2-wire resistance measurement.
34. [SENSe:]RESistance:APERture:ENABled
Command Format:
[SENSe:]RESistance:APERture:ENABled□{ON|OFF}
[SENSe:]RESistance:APERture:ENABled?
Function:
The command sets or queries the enable state of 2WR interval time measuring
function.
Explanations:
[SENSe:]RESistance:APERture:ENABled □ {ON|OFF}: Enable or disable 2WR
interval time measuring function.
[SENSe:]RESistance:APERture:ENABled?: Query the enable state of 2WR interval
time measuring function.
The commands ―CONFigure:RESistance‖, ―MEASure:RESistance?‖, ―[SENSe:]
RESistance:NPLC‖ and ―[SENSe:]RESistance:RESolution‖ can be used to disable
interval time measuring function (OFF).
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.