RIGOL
Programming Guide for DM3058
Command Format:
[SENSe:]FREQuency:NULL[:STATe]□{ON|OFF}
[SENSe:]FREQuency:NULL[:STATe]?
Function:
The command sets or queries the enable state of frequency NULL measuring
function.
Explanations:
[SENSe:]FREQuency:NULL[:STATe] □ {ON|OFF}: Enable or disable frequency
NULL measuring function.
[SENSe:]FREQuency:NULL[:STATe]?: Query the enable state of frequency NULL
measuring function.
The command ―[SENSe:]FREQuency:NULL:VALue‖ can be used to set NULL
value.
The commands ―CONFigure:FREQuency‖ and ―MEASure:FREQuency?‖ can be
used to disable NULL function (OFF).
Return Format:
The query returns the enable state of frequency NULL measuring function.
54. [SENSe:]FREQuency:NULL:VALue
Command Format:
[SENSe:]FREQuency:NULL:VALue□{<value>|MIN|MAX}
[SENSe:]FREQuency:NULL:VALue?□[{MIN|MAX}]
Function:
The command sets or queries the NULL value of frequency measurement.
Parameters:
Explanations:
[SENSe:]FREQuency:NULL:VALue□{<value>|MIN|MAX}: Set the NULL value of
frequency measurement. MIN is -120% max range of current function, MAX is