Programming Guide for DM3058
22. :measure:fresistance:range?
Command Format:
:measure:fresistance:range?
Function:
The command queries the measuring range of 4-wire resistance.
Return Format:
The query returns the parameter related to current range : 0, 1, 2, 3, 4, 5 or 6
(see Table 3-5).
23. :measure:frequency?
Command Format:
:measure:frequency?
Function:
The command queries frequency.
Explanation:
The range of frequency is 20 Hz~1 MHz. This command is invalide for Dual
Display
Return Format:
The query returns the current frequency in the form of 8.485240e-05, and unit is
Hz.
24. :measure:frequency
Command Format:
:measure:frequency{<range>|MIN|MAX|DEF}
Function:
The command sets the measuring range of input voltage in frequency
measurement.