Rigol DM3058 Specifikace Strana 129

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 264
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 128
RIGOL
Programming Guide for DM3058
5-13
3. CONFigure:CONTinuity
Command Format:
CONFigure:CONTinuity
Function:
The command resets all the continuity measuring parameters and trigger
parameter as defaults, and sets continuity measuring function.
Explanations:
Enable continuity test, and set the range as 2 kΩ.
4. CONFigure:CURRent:AC
Command Format:
CONFigure:CURRent:AC
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the AC current measuring parameters and trigger
parameters as defaults, and sets AC current measuring function.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{ 20 mA |200 mA | 2 A | 10 A }
200 mA
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 20 mA (MIN), 200 mA, 2 A, 10 A (MAX), AUTO (autorange) (unit
is A). MAX = 10, MIN = 0. 02.
<resolution> (unit is A); in default setting, the rate is medium.
5. CONFigure:CURRent[:DC]
Command Format:
CONFigure:CURRent[:DC]
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Zobrazit stránku 128