Rigol DM3058 Uživatelský manuál Strana 98

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 97
RIGOL
Programming Guide for DM3058/DM3058E
4-5
CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum}
CALCulate:LIMit:UPPer? [MINimum|MAXimum]
Function:
The commands set and query the upper limit of current measurement.
Parameter:
Name Type Range Default
<value>
Consecutive
Integer
±120% of MAX range 0
Explanations:
The command is valid only when the LIMIT function (CALCulate:FUNCtion LIMIT)
is enabled under calculation is active (CALCulate:STATe ON).
CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum} - Sets the desired upper
limit under current measurement.
CALCulate:LIMit:UPPer? [MINimum|MAXimum] - Queries the upper limit under
current measurement.
MIN denotes the minimum range (-120% of MAX range) of current measurement.
MAX denotes the maximum range (120% of MAX range) of current measurement.
Return Format:
The query returns the upper limit of current measurement.
5. CALCulate:DB:REFerence
Command Format:
CALCulate:DB:REFerence {<value>|MINimum|MAXimum}
CALCulate:DB:REFerence? [MINimum|MAXimum]
Function:
The command commands set and query the dB reference value.
Parameter:
Name Type Range Default
<value> Consecutive Integer -120 dB~+120 dB 0
Zobrazit stránku 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 191 192

Komentáře k této Příručce

Žádné komentáře