Rigol DM3058 Uživatelský manuál Strana 126

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 125
RIGOL
Programming Guide for DM3058/DM3058E
4-33
[SENSe:]VOLTage:AC:RANGe? [MINimum|MAXimum] Queries the range of AC
voltage measurement.
Return Format:
The query returns the range of AC voltage measurement.
4. [SENSe:]CURRent:DC:RANGe
Command Format:
[SENSe:]CURRent:DC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum]
Function:
The commands set and query the range of DC current measurement
Parameter:
Name Type Range Default
<range> Discrete { 200 µA|2 mA|20 mA
|200 mA|2 A|10 A }
200 mA
Explanations:
[SENSe:]CURRent:DC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of DC current measurement from {200 µA|2 mA|20 mA|200 mA|1 A|10 A};
thereinto, MIN = 200 µA, MAX = 10 A and the unit is A. If a 200 mA range is
required, enter 0.2 as its parameter.
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum] Queries the range of DC
current measurement.
Return Format:
The query returns the range of DC current measurement.
5. [SENSe:]CURRent:AC:RANGe
Command Format:
[SENSe:]CURRent:AC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]CURRent:AC:RANGe? [MINimum|MAXimum]
Zobrazit stránku 125