Programming Guide for DM3058/DM3058E
4-53
1. TRIGger:COUNt
Command Format:
TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite}
TRIGger:COUNt? [MINimum|MAXimum]
Function:
The commands set and query the trigger times used in multimeter.
Parameter:
Name Type Range of Values Default
<value>
Consecutive
Integer
1~2000
Parameters must
be set
Explanations:
TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite} - Sets the trigger times
to be used.
TRIGger:COUNt? [MINimum|MAXimum] - Queries the current trigger times.
Return Format:
The query returns the trigger times used in multimeter.
2. TRIGger:DELay
Command Format:
TRIGger:DELay {<seconds>|MINimum|MAXimum}
TRIGger:DELay? [MINimum|MAXimum]
Function:
The commands set and query the delay time from when the trigger signal
occured until the start of the measurement.
Parameter:
Name Type Range of Values Default
<seconds>
Consecutive
Integer
0~3600
Parameters must
be set
Explanations: