RIGOL
Programming Guide for DM3058/DM3058E
3-49
The query returns the parameter related to current 2-wire resistance rate (see
Table 3-10).
Example:
Sets the rate under 2-wire resistance measurement to Medium.
:RATE:RESistance M
6. :RATE:FRESistance
Command Format:
:RATE:FRESistance?
:RATE:FRESistance {F|M|S}
Function:
The commands query and set the rate of 4-wire resistance measurement.
Explanations:
:RATE:FRESistance? - Queries the current measuring rate of 4-wire resistance.
The command is valid only when 4-wire resistance function is enabled.
:RATE:FRESistance {F|M|S} - Sets the desired measuring rate of 4-wire
resistance. For the range and the explanation of each parameter, see Table 3-10.
Return Format:
The query returns the parameter related to current 4-wire resistance rate (see
Table 3-10).
Example:
Sets the rate under 4-wire resistance measurement to Medium.
:RATE:FRESistance M
7. :RATE:SENSor
Command Format:
:RATE:SENSor?
:RATE:SENSor {M|S}
Function: