Rigol DM3058 Uživatelský manuál Strana 83

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 82
RIGOL
Programming Guide for DM3058/DM3058E
3-58
In Medium measurement, the default interval is 50 ms and the available range is
within 50 ms and 2000 ms;
In Slow measurement, the default interval is 400 ms and the available range is
within 400 ms and 2000 ms.
The default measurement rate is Slow, so the default interval of the multimeter is
400 ms.
Return Format:
The query returns the current interval of auto trigger (the unit is ms).
3. :TRIGger:AUTO:HOLD
Command Format:
:TRIGger:AUTO:HOLD?
:TRIGger:AUTO:HOLD {ON|OFF|1|0}
Function:
The commands query and set the hold funtion under auto trigger.
Explanations:
:TRIGger:AUTO:HOLD? – Queries if the hold function under auto trigger is
enabled.
:TRIGger:AUTO:HOLD {ON|OFF|1|0} Sets the desired hold state under auto
trigger.
Return Format:
The query returns the ON (1) or OFF (0).
4. :TRIGger:AUTO:HOLD:SENSitivity
Command Format:
:TRIGger:AUTO:HOLD:SENSitivity?
:TRIGger:AUTO:HOLD:SENSitivity {<value>|MIN|MAX|DEF}
Function:
The commands query and set the delay sensitive of auto trigger.