Rigol DM3058 Uživatelský manuál Strana 106

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 105
RIGOL
Programming Guide for DM3058/DM3058E
4-13
4. CONFigure:CURRent:AC
Command Format:
CONFigure:CURRent:AC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution for AC
current measurement, but does not start the measurement.
Parameter:
Name Type Range Default
<range> Discrete {20 mA|200 mA|2 A|10 A
|AUTO}
AUTO
<resolution> Discrete Refer to Table 4-2 1ppm x range
Explanations:
The unit is A.
For <range>: MIN = 0. 02 (20mA); MAX = 10, DEF=AUTO.
<resolution> relates to both the measurement rate and integration time, see
Table 4-2.
5. CONFigure:RESistance
Command Format:
CONFigure:RESistance
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution for 2-wire
resistance measurement, but does not start the measurement.
Parameter:
Name Type Range Default
<range> Discrete {200 Ω|2 kΩ|20 kΩ|200 kΩ
|2 MΩ|10 MΩ|100 MΩ|AUTO}
AUTO
<resolution> Discrete Refer to Table 4-2 1ppm x range
Zobrazit stránku 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 191 192

Komentáře k této Příručce

Žádné komentáře