Rigol DM3058 Uživatelský manuál Strana 57

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 56
RIGOL
Programming Guide for DM3058/DM3058E
3-32
The command queries the current range of AC voltage measurement.
Return Format:
The query returns: 0, 1, 2, 3 or 4 (see Table 3-5).
11. :MEASure:CURRent:DC?
Command Format:
:MEASure:CURRent:DC?
Function:
The command queries the measured DC current value.
Explanations:
The command is unavailable in double-display.
Return Format:
The command queries the measured DC current value in scientific notation, such
as: 9.67441e-05, the unit is A.
12. :MEASure:CURRent:DC
Command Format:
:MEASure:CURRent:DC {<range>|MIN|MAX|DEF}
Function:
The command sets the range and resolution of DC current measurement.
Parameter:
Name Type Range Default
< range> Discrete {0|1|2|3|4|5} 0
Explanations:
The corresponding resultion will be set automatically after you set the range.
The measurement type will change into Manualautomaticlly once you set the
range.
For the parameter range and the corresponding resolution, see table below.
Zobrazit stránku 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 191 192

Komentáře k této Příručce

Žádné komentáře