RIGOL Chapter 2 Command System
Programming Guide for DM3000
<Range> can be AUTO|0|1|2|3|4|5|6, AUTO denotes using
auto range measurement, for more details about parameter 0 to
6, please refer to the “Explanation” in :MEASure commands.
<Resolution>: denotes the measurement reading resolution of
the current task and can be 0|1|2.
<SalmpleNum> denotes the number of samples specified by the
current task and ranges from 1 to 100.
<Delay> denotes the interval among samples of the current
task and ranges from 0 to 360000, the default unit is s.
:SCAN:TASK:ADD 2,5,DCV,2,1,25,10
:SCAN:TASK:DELEte <TasNum>
Deletes the task specified by <TasNum>.
<TasNum> ranges from 0 to the maximum task number in current
scan project.
:SCAN:TASK:INTErval <time>
Sets the tasks interval under a scan project, the default unit is s.
<time> ranges from 0 to 3600.
The query returns the created scan task information such as:
00:CH05,DCV,3,1,25;01:CH05,DCV,3,1,25;02:CH05,DCV,3,1,25;
If no tasks are available under the current scan project, the query
returns NULL.
Queries the running state of the current scan task.
The query returns RUN or STOP.
Sets the number of cycles for a scan task and executes this task.