Chapter 2 Command System RIGOL
Programming Guide for DM3000
The format of
<ip_address>
is “nnn.nnn.nnn.nnn” and all
“nnn” range from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto Ip should be disabled while you use this
command.
6. :UTILity:INTErface:LAN:GATEway
:UTILity:INTErface:LAN:GATEway?
:UTILity:INTErface:LAN:GATEway
<ip_address>
Defines the gate way of the network that currently connected to
the meter.
The format of
<ip_address>
is “nnn.nnn.nnn.nnn”. The first
“nnn” ranges from 0 to 223 (except 127) and the others range
from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto IP should be disabled while you use this
command.
7. :UTILity:INTErface:LAN:DNS
:UTILity:INTErface:LAN:DNS?
:UTILity:INTErface:LAN:DNS
<ip_address>
Defines the gate DNS server address of the network that currently
connected to the meter.
The format of
<ip_address>
is “nnn.nnn.nnn.nnn”. The first
“nnn” ranges from 0 to 223 (except 127) and the others range
from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto IP should be disabled while you use this
comand.
8. :UTILity:INTErface:GPIB:ADDRess
:UTILity:INTErface:GPIB:ADDRess?
:UTILity:INTErface:GPIB:ADDRess
<value>
Sets the GPIB address of the meter.