Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-33
Explanation
The format of
<ip_address>
is nnn.nnn.nnn.nnn and all
nnn range from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto Ip should be disabled while you use this
command.
Default
255.255.255.0
6. :UTILity:INTErface:LAN:GATEway
Syntax
:UTILity:INTErface:LAN:GATEway?
:UTILity:INTErface:LAN:GATEway
<ip_address>
Function
Defines the gate way of the network that currently connected to
the meter.
Explanation
The format of
<ip_address>
is nnn.nnn.nnn.nnn. The first
nnn ranges from 0 to 223 (except 127) and the others range
from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto IP should be disabled while you use this
command.
Default
172.16.3.1
7. :UTILity:INTErface:LAN:DNS
Syntax
:UTILity:INTErface:LAN:DNS?
:UTILity:INTErface:LAN:DNS
<ip_address>
Function
Defines the gate DNS server address of the network that currently
connected to the meter.
Explanation
The format of
<ip_address>
is nnn.nnn.nnn.nnn. The first
nnn ranges from 0 to 223 (except 127) and the others range
from 0 to 255.
The IP address configuration type should be Manual and both
DHCP and Auto IP should be disabled while you use this
comand.
Default
0.0.0.0
8. :UTILity:INTErface:GPIB:ADDRess
Syntax
:UTILity:INTErface:GPIB:ADDRess?
:UTILity:INTErface:GPIB:ADDRess
<value>
Function
Sets the GPIB address of the meter.