RIGOL Chapter 2 Command System
Programming Guide for DM3000
secondary screen and the meter is measuring AC voltage.
18. :MEASure:VOLTage:AC:FREQuency:STATe?
:MEASure:VOLTage:AC:FREQuency:STATe?
Queries whether the frequency was displayed on the secondary
screen while measuring AC voltage.
The query returns DISPLAY or HIDE.
The query returns the DC current measurement value in the form of
scientific notation such as -3.74725404E-06, the unit is A.
:MEASure:CURRent:DC {0|1|2|3|4|MIN|MAX|DEF}
Sets the measurement range of DC current.
The measurement mode will change to “Manual” while you set
the range.
Each parameter has its own range:
Setting the measurement range of DC current to maximum:
:MEASure:CURRent:DC 4 or
:MEASure:CURRent:DC MAX
21. :MEASure:CURRent:DC:RANGe?
:MEASure:CURRent:DC:RANGe?
Queries the measurement range of DC current.
The query returns 0, 1, 2, 3 or 4.