Rigol DS2000A Series Specifikace Strana 336

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 390
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 335
RIGOL 2 Command System
2-320 MSO2000A/DS2000A Programming Guide
:TRIGger:SHOLd:TYPe
Syntax
:TRIGger:SHOLd:TYPe <type>
:TRIGger:SHOLd:TYPe?
Description
Set or query the hold type of setup/hold trigger.
Parameter
Name Type Range Default
<type> Discrete {SETup|HOLd|SETHOLd} SETup
Explanation
SETup: the oscilloscope triggers when the setup time is less than the setting value (refer to
the
:TRIGger:SHOLd:STIMe command).
HOLd: the oscilloscope triggers when the hold time is less than the setting value (refer to
the
:TRIGger:SHOLd:HTIMe command).
SETHOLd: the oscilloscope triggers when the setup time or hold time is less than the corresponding
setting value (refer to the
:TRIGger:SHOLd:STIMe and :TRIGger:SHOLd:HTIMe commands).
Return Format
The query returns SET, HOL or SETHOL.
Zobrazit stránku 335
1 2 ... 331 332 333 334 335 336 337 338 339 340 341 ... 389 390

Komentáře k této Příručce

Žádné komentáře