RIGOL 2 Command System
2-290 MSO2000A/DS2000A Programming Guide
:TRIGger:EDGe
:TRIGger:EDGe:LEVel
:TRIGger:EDGe:SLOPe
:TRIGger:EDGe:SOURce
:TRIGger:EDGe:LEVel
Syntax
:TRIGger:EDGe:LEVel <level>
:TRIGger:EDGe:LEVel?
Description
Set or query the trigger level in edge trigger when the trigger source is CHANnel1, CHANnel2 or EXT. The
unit is the same with the current amplitude unit.
Parameter
Name Type Range Default
<level> Real
When the trigger source is CHANnel1 or CHANnel2:
-5×VerticalScaleOFFSet to +5×VerticalScale–OFFSet
When the trigger source is EXT: -4V to 4V
0
Note: For the VerticalScale, refer to the :CHANnel<n>:SCALe command. For the OFFSet, refer to
the
:CHANnel<n>:OFFSet command.
Explanation
To set the trigger source of edge trigger, please refer to the
:TRIGger:EDGe:SOURce command.
When the trigger source of edge trigger is a digital channel, please use the
:LA:POD<n>:THReshold
command to set or query the threshold level of edge trigger.
When the trigger source of edge trigger is ACLine, you do not need to set this parameter.
Return Format
The query returns the trigger level in edge trigger when the trigger source is CHANnel1, CHANnel2 or EXT
in scientific notation.