RIGOL 2 Command System
2-270 MSO2000A/DS2000A Programming Guide
[:TRACe<n>]:DATA:LOAD?
Syntax
[:TRACe<n>]:DATA:LOAD? 1
Description
Read the specified data packet in the volatile memory of the specified signal source.
Parameter
Name Type Range Default
<n> Discrete {1|2} 1
Return Format
The query returns a binary data block. The data block header is #9000032768 followed by 32768 bytes
binary data.
[:TRACe<n>]:DATA:POINts
Syntax
[:TRACe<n>]:DATA:POINts VOLATILE, {<point>|MINimum|MAXimum}
[:TRACe<n>]:DATA:POINts? VOLATILE[,MINimum|MAXimum]
Description
Set or query the number of initial points of the arbitrary waveform of the specified signal source channel.
Parameter
Name Type Range Default
<point> Integer 1 to 16384 2
<n> Discrete {1|2} 1
Explanation
This command sets the amplitudes of the initial points of the volatile waveform to 0μV automatically.
After sending this command, you can use the
[:TRACe<n>]:DATA:VALue command to modify the
voltage of the specified point.
Return Format
The query returns an integer between 1 and 16384.
Example
:DATA:POINts VOLATILE,200