2-31
X-Y Format
This format is useful for studying phase relationships between CH1 and ch2 (CH3
and CH4 cant be used).CH1 in the horizontal axis(X) and CH2 in the vertical axis(Y),
the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Press MENUTime BaseX-Y to enter X-Y mode.
Figure 2-37 X-Y display format
The oscilloscope can acquire waveforms with any sampling rate in Y-T mode (up to
2GSa/s for single channel or half channel
[1]
). But in X-Y mode, up to 1GSa/s can be
used and the default is 250MSa/s. Generally, the lower the sample rate is, the better
the Lissajous graphic will be.
NOTE [1]: Half channel indicates selecting one of the channels in CH1 and CH2, or in CH3 and
CH4.
The following modes or functions will not work in X-Y format.
Auto cursor measure mode
Track cursor measure mode
REF and MATH Operations
Delayed Scan Mode
Vector Display Mode
Horizontal knob
Trigger Controls(excluding MODE button)