Delta Time
Measurement Accuracy
(Full Bandwidth)
Single-shot: ± (1 sample interval + 50ppm × reading +
0.6 ns)
>16 averages: ± (1sample interval + 50ppm × reading
+ 0.4 ns)
8-bit resolution, each channel samples simultaneously
2mV/div-10V/div at input BNC
± 40V(250mV/div ~10V/div)
± 2V(2mV/div~245mV/div)
60MHz(DS1062/4B)
100MHz(DS1102/4B)
200MHz(DS1202/4B)
60MHz(DS1062/4B)
100MHz(DS1102/4B)
200MHz(DS1202/4B)
Selectable Analog
Bandwidth Limit
Lower Frequency
Limit (AC -3dB)
Rise Time at BNC,
typical
<1.75ns, <3.5ns, <5.8ns,
On 200MHz, 100MHz, 60MHz respectively
2mV/div~5mV/div: ± 4% (Normal or Average acquisition
mode)
10mV/div~10V/div: ± 3% (Normal or Average acquisition
mode)
DC Measurement
Accuracy, Average
Acquisition Mode
Average of ≥16 Waveforms with vertical position at zero:
±(DC Gain Accuracy× reading+0.1div+1mV)
Average of ≥16 Waveforms with vertical position not at zero:
±[DC Gain Accuracy× (reading+ vertical position)+(1% of
vertical position)+0.2div]
Add 2mV for settings from 2mV/div to 245 mV/div
Add 50mV for settings from 250mV/div to 10V/div
Delta Volts
Measurement
Accuracy (Average
Delta Volts between any two averages of 16 waveforms
acquired under same setup and ambient
conditions: ±(DC Gain Accuracy× reading + 0.05 div)