RIGOL Chapter 10 Pass/Fail Test
10-4 MSO2000A/DS2000A Users Guide
waveforms are detected. The test results on the screen will update
continuously and a pulse will be output from the [Trigger Out] connector
at the rear panel each time a failed waveform is detected.
3. Beeper on Fail
Press Output to select “Fail” or “Fail+ ”.
Fail: when a failed waveform is detected, the beeper does not sound.
Fail+
: when a failed waveform is detected, the beeper sounds (not
related to the on/off state of the sound).
4. [Trigger Out] Outputs Pulse
Press AuxOutput to selectONor OFF. If ONis selected, the oscilloscope
outputs a pulse from the [Trigger Out] connector at the rear panel when a
failed waveform is detected.
You can also press Utility AuxOutput and select PassFailto set this
output.
To Save the Test Mask
Users can save the current test mask in *.pfformat to the internal Flash memory or
external USB storage device. The internal memory can store at most 10 test mask
files (LocalPF.pf).
Press Save to enter the file store interface. Please refer to the relative descriptions in
Store and Recallto save the test mask file to the internal or external memory.
To Load the Test Mask
Users can also load the test mask files (*.pf) stored in the internal Flash memory or
external USB storage device to the instrument.
Press Load to enter the file recall interface. Please refer to the relative descriptions
in
Store and Recallto load the test masks to the instrument.