RIGOL User’s Guide DS1000B Series Digital Oscilloscopes DS1074B,DS1104B,DS1204B Dec. 2014 RIGOL Technologies, Inc.

Strany 2

RIGOL User’s Guide for DS1000B Series VIII Contents Guaranty and Declaration ...

Strany 3 - Guaranty and Declaration

RIGOL User’s Guide for DS1000B series 2-70 2. Remote Control (1) By Web Page When the network configuration is completed, open the IE and enter t

Strany 4 - Safety Notices

RIGOL User’s Guide for DS1000B Series 2-71 (2) By Commands DS1000B is a USB TMC (USB Test & Measurement Class) instrument, which could be contro

Strany 5

RIGOL User’s Guide for DS1000B series 2-72 Figure 2-93 VISA TCP/IP Resource interface  Click “Next” and select “Auto-detect of LAN Instrument”.

Strany 6 - Safety Terms and Symbols

RIGOL User’s Guide for DS1000B Series 2-73  The software searching the available network device automatically. Figure 2-95 Search the network dev

Strany 7

RIGOL User’s Guide for DS1000B series 2-74 Figure 2-97 Communication testing dialog  Make a connection by USB  Connect the oscilloscope to a

Strany 8

RIGOL User’s Guide for DS1000B Series 2-75 Figure 2-98 Measurement & Automation Explorer interface  Select the responding resource and open t

Strany 9 - Document Overview

RIGOL User’s Guide for DS1000B series 2-76 Language The DS1000B series oscilloscopes have multi-language user menu, choose as your desire. Press Ut

Strany 10 - Contents

RIGOL User’s Guide for DS1000B Series 2-77 Pass/Fail The Pass/Fail function monitors changes of signals by comparing the input signal which is within

Strany 11

RIGOL User’s Guide for DS1000B series 2-78 1. Mask setting Press Utility→Pass/Fail→Mask Setting to go to the following menu. Figure 2-103 Table

Strany 12

RIGOL User’s Guide for DS1000B Series 2-79 Figure 2-105 Table 2-59 The Mask setting menu (External, Page 2/2) Menu Settings Comments Save Go to

Strany 13 - Chapter 1 Quick Guide

RIGOL User’s Guide for DS1000B Series IX Waveform Recorder ... 2-80 Print Setting .

Strany 14 - General Inspect

RIGOL User’s Guide for DS1000B series 2-80 Waveform Recorder Waveform recorder records input waveforms from CH1, CH2, CH3 and CH4 with a maximum rec

Strany 15 - Panels and User Interface

RIGOL User’s Guide for DS1000B Series 2-81 Figure 2-108 Table 2-62 The Record menu (Page 2/2) Menu Settings Comments Interval <1.00ms-1000s>

Strany 16 - Rear Panel

RIGOL User’s Guide for DS1000B series 2-82 Run/Stop Button The Run/Stop button can also replay or continue the waveform display. The Current Frame w

Strany 17

RIGOL User’s Guide for DS1000B Series 2-83 Figure 2-113 Table 2-67 The Storage menu (External location Page 2/2) Menu Settings Comments Save Go

Strany 18 - User Interface

RIGOL User’s Guide for DS1000B series 2-84 Print Setting DS1000B series oscilloscopes support two print modes: “normal” and “PictBridge”. Please exe

Strany 19 - Functional Check

RIGOL User’s Guide for DS1000B Series 2-85 2. To set up print parameters: Press Utility→Print set to into a setting menu. Please setup the printer

Strany 20

RIGOL User’s Guide for DS1000B series 2-86 Figure 2-118 Table 2-70 Print setup menu (PictBridge mode, Page 2/4) Menu Settings Comments Paper si

Strany 21 - To Compensate the Probes

RIGOL User’s Guide for DS1000B Series 2-87 3. Print: Make sure the Printer has been connected and the print settings are all ready before printing

Strany 22

RIGOL User’s Guide for DS1000B series 2-88 Self-Calibration The Self-Calibration adjusts the internal circuitry to get the best accuracy. Use this f

Strany 23

RIGOL User’s Guide for DS1000B Series 2-89 System Information Press Utility→System Info to view the system information, including Model, Serial N

Strany 25

RIGOL User’s Guide for DS1000B series 2-90 Preference Press Utility→Preference to go to the following menu Figure 2-123 Table 2-73 The Preference

Strany 26 - ) to the zero point

RIGOL User’s Guide for DS1000B Series 2-91 Date and Time DS1000B Series provide build-in system real-time date and time. Users can set the date and t

Strany 27

RIGOL User’s Guide for DS1000B series 2-92 To Measure Automatically The Measure button in the menu area activates the automatic measurement functi

Strany 28

RIGOL User’s Guide for DS1000B Series 2-93 Figure 2-128 Table 2-76 The Measure menu (Page 2/2) Menu Settings Comments Display All OFF ON Turn off al

Strany 29

RIGOL User’s Guide for DS1000B series 2-94 Quick Measurement Setup Press Measure→QuickMea button, go to the quick measure menu。 Figure 2-129 Table

Strany 30

RIGOL User’s Guide for DS1000B Series 2-95 Delay/Phase Setup Press Measure→Delay/Phase Setup button, and go to the delay/phase setup menu. Figure 2-

Strany 31

RIGOL User’s Guide for DS1000B series 2-96 Voltage Measurements The DS1000B series oscilloscopes provide automatic voltage measurements including Vp

Strany 32

RIGOL User’s Guide for DS1000B Series 2-97 Press Measure→Voltage button, and go to the following menu. Figure 2-132 Table 2-79 The Voltage Measurem

Strany 33

RIGOL User’s Guide for DS1000B series 2-98 Time Measurements The DS1000B series oscilloscopes provide 12 kinds of timing parameters auto-measure

Strany 34 - Channel Settings

RIGOL User’s Guide for DS1000B Series 2-99 Press Measure→Time button, and go to the following menu Figure 2-136 Table 2-82 The Time Measurement me

Strany 35

RIGOL User’s Guide for DS1000B Series 1-1 Chapter 1 Quick Guide This chapter covers the following topics:  General Inspect  Panels and User I

Strany 36

RIGOL User’s Guide for DS1000B series 2-100 Key Points: 1. Select the signal channel for measuring. CH1, CH2, CH3, CH4 according to the interest.

Strany 37

RIGOL User’s Guide for DS1000B Series 2-101 To Measure with Cursors Figure 2-140 shows the Cursor button on the front-panel. Figure 2-140 The Cur

Strany 38

RIGOL User’s Guide for DS1000B series 2-102 Manual Mode Press Cursor→Mode→Manual to display the manual mode menu. Figure 2-141 Table 2-85 The Curs

Strany 39

RIGOL User’s Guide for DS1000B Series 2-103 Table 2-86 The Cursor menu Cursor Increment Operation Cursor A X Y Turn the multifunctional knob () to

Strany 40

RIGOL User’s Guide for DS1000B series 2-104 Track Mode Press Cursor→Mode→Track to display the manual mode menu. Figure 2-142 Table 2-87 Cursor des

Strany 41

RIGOL User’s Guide for DS1000B Series 2-105 Table 2-88 The Cursor usage Cursor Operation Cursor A Turn the multifunctional knob ( ) to move cursor

Strany 42

RIGOL User’s Guide for DS1000B series 2-106 Auto mode Press Cursor→Mode→Auto to display the manual mode menu. Figure 2-143 Table 2-89 The Mode set

Strany 43

RIGOL User’s Guide for DS1000B Series 2-107 To Use Run Control Buttons The RUN control buttons include Auto (auto setting), Run/Stop and Single.

Strany 44 - Math Function

RIGOL User’s Guide for DS1000B series 2-108 Auto-Set functions When the Auto is pressed the oscilloscope is configured to the following defaults: Ta

Strany 45

RIGOL User’s Guide for DS1000B Series 3-1 Chapter 3 Application Examples Example 1: Taking Simple Measurements The function is used to observe an

Strany 46

RIGOL User’s Guide for DS1000B series 1-2 General Inspect After receiving a new DS1000B series oscilloscope, please inspect the instrument as foll

Strany 47

RIGOL User’s Guide for DS1000B Series 3-2 Example 2: View a Signal Delay Caused by a Circuit Th

Strany 48

RIGOL User’s Guide for DS1000B Series 3-3 Example 3: Capture a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowl

Strany 49

RIGOL User’s Guide for DS1000B Series 3-4 Example 4: To Reduce the Random Noise on a Signal If

Strany 50 - REF Function

RIGOL User’s Guide for DS1000B Series 3-5 4. To reduce the noise by setting the acquisition type and adjust the waveform intensity. (1) If there is

Strany 51

RIGOL User’s Guide for DS1000B Series 3-6 Example 5: Making Cursor Measurements There are 22 b

Strany 52

RIGOL User’s Guide for DS1000B Series 3-7 Measure the Amplitude of the First Waveform Peak of the Sinc. Please follow these steps: 1. Press Cursor k

Strany 53

RIGOL User’s Guide for DS1000B Series 3-8 Example 6: The Application of the X-Y operation View

Strany 54

RIGOL User’s Guide for DS1000B Series 3-9 Figure 3-6 Ellipse method to observe the phase difference Sinθ = A/B or C/D, where θ = phase shift (in de

Strany 55

RIGOL User’s Guide for DS1000B Series 3-10 Example 7: Triggering on a Video Signal Test a vide

Strany 56

RIGOL User’s Guide for DS1000B Series 3-11 Triggering on the Video Lines 1. Press the MENU key in the Trigger control area to see the trigger menu.

Strany 57

RIGOL User’s Guide for DS1000B Series 1-3 Panels and User Interface Being familiar with the operation panels is the first priority when you get a

Strany 58

RIGOL User’s Guide for DS1000B Series 3-12 Example 8: FFT Cursor Measurement FFT measurements

Strany 59

RIGOL User’s Guide for DS1000B Series 3-13 Example 9: Pass/Fail Test In Pass/Fail Test, the oscilloscope compares the input signal to the establis

Strany 61 - Delayed Scan

RIGOL User’s Guide for DS1000B Series 4-1 Chapter 4 Troubleshooting 1. After the oscilloscopes is powered on, th

Strany 62 - X-Y Format

RIGOL User’s Guide for DS1000B Series 4-2 the Trigger Mode as “AUTO”. Moreover, push Auto button

Strany 63 - To Set up the Trigger System

RIGOL User’s Guide for DS1000B Series 5-1 Chapter 5 Specifications All specifications apply to the DS1000B Series

Strany 64 - Edge Trigger

RIGOL User’s Guide for DS1000B Series 5-2 Technical Specifications Acquisition Sampling Modes R

Strany 65 - Pulse Width Trigger

RIGOL User’s Guide for DS1000B Series 5-3 Sample Rate and Delay Time Accuracy ±50ppm (over any≥1ms time interval) Ver

Strany 66 - Video Trigger

RIGOL User’s Guide for DS1000B Series 5-4 Trigger Trigger Sensitivity 0.1div-1.0div (adjustable

Strany 67 - < Line sync >

RIGOL User’s Guide for DS1000B Series 5-5 Reciprocal of ∆T in Hertz (1/∆T) Track Voltage value for Y-axis waveform Ti

Strany 68

RIGOL User’s Guide for DS1000B series 1-4 Front Panel Interfaces: USB Host: Connect DS1000B with external USB devices when the oscilloscope is regar

Strany 69 - Pattern Trigger

RIGOL User’s Guide for DS1000B Series 5-6 General Specifications Display Display Type 5.7 inch.

Strany 70

RIGOL User’s Guide for DS1000B Series 6-1 Chap

Strany 71 - Alternative Trigger

RIGOL User’s Guide for DS1000B Series 6-2 Appendix B: Warranty RIGOL warrants its products’ mainframe and accessories in materials and workmanship

Strany 72

RIGOL User’s Guide for DS1000B Series 6-3 Appe

Strany 74 - Trigger Settings

RIGOL User’s Guide for DS1000B Series 1 Index

Strany 75

RIGOL . User’s Guide for DS1000B Series 2 UTILITY ... 2-64 Vertical ... 5-3 Vertic

Strany 76 - Trigger Holdoff

RIGOL User’s Guide for DS1000B Series 1-5 Notation definitions in this Manual: Throughout this guide, notation symbols of buttons and knobs are the

Strany 77 - Trigger Terms Explanation

RIGOL User’s Guide for DS1000B series 1-6 User Interface Figure 1-3 User Inte

Strany 78

RIGOL User’s Guide for DS1000B Series 1-7 Functional Check Perform this quick functional check to verify that the instrument is operating correct

Strany 80

RIGOL User’s Guide for DS1000B series 1-8 Figure 1-5 Probe connection Figure 1-6 Probe adjustment [1]  Attach the probe tip to the Pr

Strany 81

RIGOL User’s Guide for DS1000B Series 1-9 To Compensate the Probes Perform this adjustment to match the characteristics of the probe and the channe

Strany 82

RIGOL User’s Guide for DS1000B series 1-10 To Display a Signal Automatically The oscilloscope has an automated feature to display the input signal

Strany 83

RIGOL User’s Guide for DS1000B Series 1-11 To Understand the Vertical System As shown in Figure 1-8, there are several buttons and knobs in the VE

Strany 84 - To Set up the Display System

RIGOL User’s Guide for DS1000B series 1-12 Measurement hints If the channel is DC coupled, measuring the DC components of the signal by simply notin

Strany 85

RIGOL User’s Guide for DS1000B Series 1-13 To Understand the Horizontal System Figure1-9 shows the HORIZONTAL controls: MENU button, and kn

Strany 86

RIGOL User’s Guide for DS1000B series 1-14 3. The horizontal knob moves displayed signal horizontally on the waveform window. The horizontal knob

Strany 87 - To Store and Recall

RIGOL User’s Guide for DS1000B Series 1-15 To Understand the Trigger System Figure 1-10 shows the TRIGGER control: MODE, MENU, FORCE, 50%

Strany 88

RIGOL User’s Guide for DS1000B series 1-16 3. Press MENU button in the Trigger control area. A soft button menu appears on the display showing the

Strany 89

RIGOL User’s Guide for DS1000B Series 1-17 To Understand the Quick Function The following figure shows that there are two buttons in the QUICK func

Strany 90 - Internal Memory

RIGOL User’s Guide for DS1000B Series I Guaranty and Declaration Copyright © 2009 RIGOL Technologies, Inc. All Rights Reserved. Trademark Inform

Strany 92 - U-disk Selection

RIGOL User’s Guide for DS1000B Series 2-1 Chapter 2 Operating Your Oscilloscope By now, a user should understand the VERTICAL, HORIZONTAL and TRI

Strany 93 - Disk Management

RIGOL User’s Guide for DS1000B series 2-2 To Set up the Vertical System Four analog channels are provided by DS1000B: CH1, CH2, CH3 and CH4. Pres

Strany 94 - Rename the file

RIGOL User’s Guide for DS1000B Series 2-3 Knobs Explanation Use the vertical controls to display signal waveforms by adjusting the vertical knob, a

Strany 95

RIGOL User’s Guide for DS1000B series 2-4 Channel Settings Each channel of DS1000B series digital oscilloscope has an independent operation menu an

Strany 96 - To Set up the Utility

RIGOL User’s Guide for DS1000B Series 2-5 Figure 2-2 Table 2-3 Channel menu (Page 2/2) Menu Settings Comments 2/2 Back to the previous menu page

Strany 97

RIGOL User’s Guide for DS1000B series 2-6 1. To set up channel coupling To use Channel 1 as an example, input a square wave signal with DC shift.

Strany 98 - The I/O Setup

RIGOL User’s Guide for DS1000B Series 2-7 Press CH1→Coupling→GND, to set “GND” coupling, it disconnects the input signal. The screen displays as Figu

Strany 99

RIGOL User’s Guide for DS1000B series 2-8 2. To set up the channel bandwidth limit Take Channel 1 for an example, input a signal containing high f

Strany 100

RIGOL User’s Guide for DS1000B Series 2-9 3. To set up Probe Attenuation The oscilloscope allows selecting the attenuation factor for the probe. Th

Strany 101

RIGOL User’s Guide for DS1000B Series II Safety Notices view the following safety precautio

Strany 102

RIGOL User’s Guide for DS1000B series 2-10 4. Digital Filter Press CH1→ Digital Filter to display the digital filter menu. Figure 2-9 Turn off d

Strany 103

RIGOL User’s Guide for DS1000B Series 2-11 Figure 2-11 Table 2-5 The Filter menu Menu Settings Comments Digital Filter ON OFF Turn on the digital

Strany 104

RIGOL User’s Guide for DS1000B series 2-12 5. To change the Volts/Div settings The Volts/Div control has Coarse or Fine configuration. The Vertica

Strany 105

RIGOL User’s Guide for DS1000B Series 2-13 6. To invert a waveform Invert turns the displayed waveform 180 degrees with respect to the ground level

Strany 106 - Language

RIGOL User’s Guide for DS1000B series 2-14 Math Function The mathematic functions include “add”, “subtract”, “Multiply” and “FFT” for CH1, CH2, CH3

Strany 107 - ■ (STOP)

RIGOL User’s Guide for DS1000B Series 2-15 Figure 2-17 Table 2-7 The Math menu (Page 2/2) Menu Settings Comments The multifunctional knob ( ) ad

Strany 108

RIGOL User’s Guide for DS1000B series 2-16 1. Using FFT function The FFT (Fast Fourier Transform) process converts a time-domain signal into its f

Strany 109

RIGOL User’s Guide for DS1000B Series 2-17 Key points for FFT 1. Signals that have a DC component or offset can cause incorrect FFT waveform compone

Strany 110 - Waveform Recorder

RIGOL User’s Guide for DS1000B series 2-18 2. Selecting an FFT Window DS1000B series oscilloscope provides four FFT windows. Each window is a tra

Strany 111 - <1.00ms-1000s>

RIGOL User’s Guide for DS1000B Series 2-19 Term Explanations FFT Resolution: The FFT resolution is the quotient between sampling rate and the number

Strany 112

RIGOL User’s Guide for DS1000B Series III interior of the device or electric shock, please do not operate in a humid environment. Do Not Operate in

Strany 113

RIGOL User’s Guide for DS1000B series 2-20 REF Function Reference Waveforms are saved waveforms to be selected for display. The reference function w

Strany 114 - Print Setting

RIGOL User’s Guide for DS1000B Series 2-21 Figure 2-21 Table 2-12 REF menu when using internal/external memory (Page 2/3) Menu Settings Comments S

Strany 115

RIGOL User’s Guide for DS1000B series 2-22 1. Import and Export Switch to external memory and select Internal location. Then, press REF →Imp. /Exp

Strany 116

RIGOL User’s Guide for DS1000B Series 2-23 2. Export Switch to external memory and select Internal location. Then, press REF→Imp. /Exp.→ Export

Strany 117

RIGOL User’s Guide for DS1000B series 2-24 3. Save to External Memory Switch to external memory and select External location. Then, press REF→Save

Strany 118 - Self-Calibration

RIGOL User’s Guide for DS1000B Series 2-25 4. New File (or New Folder) Switch to external memory and select External location. Then, press REF→S

Strany 119

RIGOL User’s Guide for DS1000B series 2-26 5. Import Switch to external memory and select External location. Then, press REF→Import and enter t

Strany 120 - Preference

RIGOL User’s Guide for DS1000B Series 2-27 6. Display a Reference Waveform Figure 2-34 Reference waveform display Operation introductions: 1.

Strany 121 - Date and Time

RIGOL User’s Guide for DS1000B series 2-28 To Set up the Horizontal System Press MENU button to enter the horizontal setting system. Figure 2-3

Strany 122 - To Measure Automatically

RIGOL User’s Guide for DS1000B Series 2-29 Term Explanations Slow Scan Mode: This mode is available when the horizontal time base is set to 50ms or sl

Strany 123

RIGOL User’s Guide for DS1000B Series IV Safety Terms and Symbols Terms in this Manual. Thes

Strany 124 - Quick Measurement Setup

RIGOL User’s Guide for DS1000B series 2-30 Knobs Explanation The horizontal controls change the horizontal scale and position of waveforms. The hori

Strany 125 - Delay/Phase Setup

RIGOL User’s Guide for DS1000B Series 2-31 Delayed Scan The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to loca

Strany 126 - Voltage Measurements

RIGOL User’s Guide for DS1000B series 2-32 X-Y Format This format is useful for studying phase relationships between CH1 and ch2 (CH3 and CH4 can’t

Strany 127

RIGOL User’s Guide for DS1000B Series 2-33 To Set up the Trigger System The trigger determines when the oscilloscope starts to acquire data and dis

Strany 128 - Time Measurements

RIGOL User’s Guide for DS1000B series 2-34 Edge Trigger An edge trigger determines whether the oscilloscope finds the trigger point on the rising or

Strany 129

RIGOL User’s Guide for DS1000B Series 2-35 Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be dete

Strany 130

RIGOL User’s Guide for DS1000B series 2-36 Video Trigger Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video si

Strany 131 - To Measure with Cursors

RIGOL User’s Guide for DS1000B Series 2-37 Figure 2-43 Table 2-25 The Video Trigger menu (Line Num, Page 2/2) Menu Settings Comments Line Num <

Strany 132 - Manual Mode

RIGOL User’s Guide for DS1000B series 2-38 Select “Line Synchronization”: Select “Filed Synchronization”: Figure 2-45

Strany 133 - ) to move

RIGOL User’s Guide for DS1000B Series 2-39 Pattern Trigger Pattern trigger recognises the trigger condition through inspecting the code preset which

Strany 134 - Track Mode

RIGOL User’s Guide for DS1000B Series V General-Purpose Oscilloscopes The introduction of this document contains three types of DS1000B series osci

Strany 135

RIGOL User’s Guide for DS1000B series 2-40 Figure 2-48 Pattern trigger: set code as falling edge

Strany 136 - Auto mode

RIGOL User’s Guide for DS1000B Series 2-41 Alternative Trigger When alternative trigger is on, the trigger sources come from two vertical channels. T

Strany 137 - To Use Run Control Buttons

RIGOL User’s Guide for DS1000B series 2-42 Figure 2-51 Table 2-30 The Alternative trigger menu (Page 1/2 Trigger Type: Pulse) Menu Settings Comme

Strany 138 - NOTE:

RIGOL User’s Guide for DS1000B Series 2-43 Figure 2-53 Table 2-32 The Alternative trigger menu (Page 1/2 Trigger Type: Video) Menu Settings Commen

Strany 139

RIGOL User’s Guide for DS1000B series 2-44 Trigger Settings Set up different trigger settings according to different trigger modes. In Edge, Pulse

Strany 140

RIGOL User’s Guide for DS1000B Series 2-45 Figure 2-56 Table 2-35 Page 2/2 Settings for Edge, Pulse and Pattern Triggers Menu Settings comments Ho

Strany 141

RIGOL User’s Guide for DS1000B series 2-46 Trigger Holdoff You can use trigger Holdoff to stabilize a complex waveform, such as a pulse sequence. Ho

Strany 142

RIGOL User’s Guide for DS1000B Series 2-47 Trigger Terms Explanation 1. Trigger Source: Trigger occurs from several sources: Input channels (CH1, CH

Strany 143

RIGOL User’s Guide for DS1000B series 2-48 display becomes stable on the screen. Any factor results in the instability of waveforms can be detected b

Strany 144

RIGOL User’s Guide for DS1000B Series 2-49 5. Adjustable trigger sensitivity To avoid the influence of noise from the physical world, and get the st

Strany 145

RIGOL User’s Guide for DS1000B Series VI  Automatic cursor tracking measurements.  Wavefor

Strany 146

RIGOL User’s Guide for DS1000B series 2-50 To Set up the Sampling System Figure 2-59 shows, the Acquire button at the MENU of the front panel.

Strany 147

RIGOL User’s Guide for DS1000B Series 2-51 The waveform displayed on the screen will change in conjunction with the setting of Acquire menu.  To re

Strany 148

RIGOL User’s Guide for DS1000B series 2-52 Figure 2-63 Adopt Normal acquisition Figure 2-64 Adopt peak detect acquisition, signal contains envel

Strany 149

RIGOL User’s Guide for DS1000B Series 2-53 Key Points: Real-time Sampling: DS1000B has Real-time sampling rate up to 2GSa/s. At the time base 20ns o

Strany 150

RIGOL User’s Guide for DS1000B series 2-54 To Set up the Display System The menu button for the display system on the front panel is shown in the f

Strany 151 - Example 9: Pass/Fail Test

RIGOL User’s Guide for DS1000B Series 2-55 Figure 2-67 Table 2-39 The Display menu (Page 2/3) Menu Settings Comments Grading ON OFF The waveform

Strany 152

RIGOL User’s Guide for DS1000B series 2-56 Term Explanations Refresh rate: It is an important performance of digital oscilloscopes. It means the

Strany 153 - Chapter 4 Troubleshooting

RIGOL User’s Guide for DS1000B Series 2-57 To Store and Recall Figure 2-69 shows the menu button for the storage system on the front panel. Figure

Strany 154

RIGOL User’s Guide for DS1000B series 2-58 Figure 2-71 Table 2-42 The Storage menu (Bitmap storage) Menu Settings Comments Storage 8-Bitmap 24-Bit

Strany 155 - Chapter 5 Specifications

RIGOL User’s Guide for DS1000B Series 2-59 Figure 2-73 Table 2-44 The Storage menu (CSV storage) Menu Settings Comments Storage CSV Create or de

Strany 156 - Technical Specifications

RIGOL User’s Guide for DS1000B Series VII Document Overview Chapter 1 Quick Start To introduce the panls and menus to help users to realize the b

Strany 157

RIGOL User’s Guide for DS1000B series 2-60 Internal Memory Press Storage→Internal to go to the following menu. Figure 2-75 Table 2-46 The Memory

Strany 158

RIGOL User’s Guide for DS1000B Series 2-61 External Memory Press Storage→ External to go to the following interface. Figure 2-76 External memory in

Strany 159 - , Phase A→B ,Phase A→B

RIGOL User’s Guide for DS1000B series 2-62 U-disk Selection Figure 2-78 U-disk selection DS1000B series have two USB Host ports on the front and r

Strany 160 - General Specifications

RIGOL User’s Guide for DS1000B Series 2-63 Disk Management Press Storage→Disk Mgr.. Then go to the following menu. Figure 2-79 Table 2-48 The Disk

Strany 161 - Chapter 6 Appendixes

RIGOL User’s Guide for DS1000B series 2-64 1. Disk Information Press Storage →Disk Mgr. →Disk info to go to the disk information interface. Fi

Strany 162 - Appendix B: Warranty

RIGOL User’s Guide for DS1000B Series 2-65 Figure 2-83 Table 2-50 The Rename menu Menu Settings Comments To move the input focus point of file n

Strany 163

RIGOL User’s Guide for DS1000B series 2-66 To Set up the Utility Figure 2-69 shows the menu button for the utility system on the front panel. Fi

Strany 164

RIGOL User’s Guide for DS1000B Series 2-67 Figure 2-86 Table 2-52 The Utility menu (Page 2/3) Menu Settings Comments Pass/Fail Setup Pass/Fail

Strany 165

RIGOL User’s Guide for DS1000B series 2-68 The I/O Setup Press Utility→IO Setting to go to the following menu. Figure 2-88 Table 2-54 The I/O Set

Strany 166

RIGOL User’s Guide for DS1000B Series 2-69 1. LAN Setting Access the oscilloscope into Network, and press LAN Set to enter the setting interface, s