Rigol DM3062 Specifikace Strana 48

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 102
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 47
RIGOL
Service Guide for DM3000 Series
3-6
Table 3-3 Zero Offset of the Standard Equipment (DM305X)
Input
Function
Quick Test
Point
Range
Error (1 year)
Open
DC Current
2.000000mA
± 1.4μA
Q
20.00000mA
± 1.6μA
200.0000 mA
± 18μA
1.000000A
± 700μA
10.00000A
[1]
± 700μA
Short
DC Voltage
Q
400.0000mV
± 32μV
4.000000V
± 240μV
40.00000V
± 2.4mV
400.0000V
± 24mV
1000.000V
[2]
± 50mV
Short
2-Wire
Ohms/
4-Wire Ohms
[3]
400.0000Ω
± 40mΩ
4.000000kΩ
± 240mΩ
40.00000 kΩ
± 2.4Ω
Q
400.0000 kΩ
± 28Ω
4.000000MΩ
± 400Ω
100.00000MΩ
± 5kΩ
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds Off after 30 seconds
On;
[2] Each additional volt over ± 500 VDC adds 0.02 mV of error;
[3] Errors in above table are corresponding to 2-wire resistance function in math Null. Without
Math Null, the error is ± 0.2 Ω additional errors in 2–wire resistance function;
[4] The reading resolution of multimeter is 5½ ;
[5] Q: Optional quick test point.
Zobrazit stránku 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 101 102

Komentáře k této Příručce

Žádné komentáře