Rigol DS2000 Series Uživatelská příručka Strana 98

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 264
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 97
RIGOL 5 To Trigger the Oscilloscope
DS2000 User’s Guide
5-22
is selected.
: only adjust the upper limit of the trigger level. During the adjustment,
“UP Level” and “Slew Rate” change accordingly but “Low Level” remains
unchanged.
: only adjust the lower limit of the trigger level. During the adjustment,
“Low Level” and “Slew Rate” change accordingly but “UP Level” remains
unchanged.
: adjust the upper and lower limits of the trigger level at the same time.
During the adjustment, “UP Level” and “Low Level” change accordingly but
“Slew Rate” remains unchanged.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Zobrazit stránku 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 263 264

Komentáře k této Příručce

Žádné komentáře