3 To Set the Horizontal System RIGOL
DS2000 User’s Guide
3-7
Note:
In Y-T mode, the oscilloscope could use any sample rate (within the guaranteed
range) to capture waveform. The maximum sample rate of X-Y mode is 1.0
GSa/s. Generally, reducing the sample rate properly could improve the display
effect of Lissajous figure.
When X-Y mode is enabled, Delayed Sweep” will be disabled automatically.
The following functions are not available in X-Y mode:
Auto measure, cursor measure, math operation, reference waveform, delayed
sweep, vector display, HORIZONTAL
POSITION, trigger control, memory
depth, acquisition mode, Pass/Fail test and waveform record.
Method 2: Use the shortcut measurement function
Please refer to Phase AB
and Phase AB measurement functions of Delay
and Phaseon page 6-17.