Rigol DS1000M Uživatelská příručka Strana 107

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 168
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 106
RIGOL
© Copyright RIGOL Technologies, Inc. 2007. 2-77
User’s Guide for DS1000 Series
Self-Calibration
The Self-Calibration adjusts the oscilloscope’s internal circuitry to get the best
accuracy. Use these functions to calibrate the oscilloscope’s vertical and horizontal
systems.
For maximum accuracy at any time, run this calibration if the ambient temperature
changes by 5° C or more.
Before running this procedure, do these steps:
1. Disconnect any probes or cables from all channel inputs, otherwise failure or
damage to the oscilloscope may occur.
2. Push the Utility button and select Self-Cal.
The self-calibration menu is displayed as Figure 2-92.
Figure 2-92
NOTE:
The oscilloscope must have been working or warm-up at least 30-minutes before
running self-calibration to get best accuracy.
Zobrazit stránku 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 167 168

Komentáře k této Příručce

Žádné komentáře