RIGOL TECHNOLOGIES USA INC
877-4-RIGOL-1 How-To Guide
www.rigolna.com
Sampling rate and Long Memory:
Using the DS1000E and DS1000D Series Rigol digital scopes, users can
access the long memory mode to get up to 512Kpts in dual channel mode
and up to 1Mpts in single channel mode. In comparison, standard memory
depth for oscilloscopes range from 1000pts - 16Kpts.
This high sampling rate is a big advantage when observing signals
where the application requires capturing a longer waveform, but also needs
to confirm higher frequency components within the signal.
The best way to visualize these higher frequency components is to use
the zoom feature. By pushing the Horizontal knob, you can enter zoom
mode. In this screen shot you see a 0.5 second waveform in yellow across
the top. In the lower section we have zoomed in to 10 usec/division. Since
we are sampling in normal mode you can see the distance between two data
points. A straight line connects data points and you can see where the signal
rises between samples and is clearly being linearly interpolated. On the
DS1000D/E Series scopes this normal mode operation has 16 Kpts per wave.
Since the overall wave is set to 50 msec/div the 10 divisions across the
screen make the wave 0.5 seconds. 16 Kpts spread across 0.5 seconds means
there is one sample approximately every 32 usec. This is confirmed by the
screen shot.