Rigol DM3064 Uživatelská příručka Strana 119

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 158
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 118
RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
3-5
Example 2: Elimination Leads Resistance Error
When measuring smaller resistance, the test leads resistance may causes large
measurement deviation.
Do these steps as follow:
1. To measure an Resistance. Connect test leads as shown below:
Figure 3- 3 Measure Resistance
2. Press
button, select Resistance (2WR) measurement function. Choose an
appropriate measurement range.
3. Set the Null measurement function parameters.
(1). Short circuit the test leads.
(2). Press Æ Null Æ Current to set the Null setting value with current
reading.
(3). Press
save this setting.
(4). Press
to finish the setting and back to a higher level menu.
4. In resistance measurement display interface, press Null, start Null function.
5. Lead test leads into circuit and start to measure.
Figure 3- 4
Resistance
Zobrazit stránku 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 157 158

Komentáře k této Příručce

Žádné komentáře