Programming Guide for DM3000 Series
2-32
Remarks:
The resistance range is a consecutive Integer among 1Ω~2000Ω.
The default resistance is 10Ω. MAX denotes the maximum is 2000Ω. MIN denotes the
minimum is . DEF denotes the default is 1.
Example:
To set the limit resistance as 1KΩ:
:measure:continuity1000
49. :measure:diode
Syntax:
:measure:diode?
Function:
The command queries diode voltage at both ends.
Remarks:
In measuring diode, the beep condition is 0.1V≤V
measured
≤2.4 V.
Return Format:
The query returns the current diode voltage in the form of 4.492510e-04, and unit is V.
(for examole)
50. :measure:diode:digit
Syntax:
:measure:diode:digit?
:measure:diode:digit{INC|DEC|5|6|7}
Function:
The command queries and sets the digit of diode voltage.
Explanations:
:measure:diode:digit? queries the digit of the diode voltage.
:measure:diode:digit<
value
> sets the digit of the diode voltage.
See Table 2- 2 .