MSO1000Z/DS1000Z Quick Guide
7
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows:
1. Perform steps 1, 2, 3 and 4 of Function Inspection.
2. Check the displayed waveforms and compare them with the following figures.
Figure 9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is as the Perfectly compensated
in the figure above.
Over compensated Perfectly compensated Under compensated