Programming Guide for DS1000/DS1000CA Series
2-76
not returned.
Example:
:TRIGger:DURation:QUALifierGREATERTHAN
Set the time condition as “Greater than”
:TRIGger:DURation:QUALifier?
Query returns GREATER THAN
8.3 :TRIGger:DURation:TIME
Command Format:
:TRIGger:DURation:TIME<time>
Explanation:
The command is to set the duration time limits. When trigger condition is being
met, trigger occurs continuously within the specified time limits.
The range of <time> is 2ns~10s.
Query Format:
:TRIGger:DURation:TIME?
Returned Format:
Query returns the set value of <time> in scientific numeric notation.
Example:
:TRIGGER:DURATION:TIME0.05
Set the limit time of the duration time as 50ms.
:TRIGGER:DURATION:TIME?
Query returns 5.00e-02.
NOTE: The Trigger system is complex, and the settings must be defined
correctly. Be patient and cautious!