User’s Guide for DS1000E, DS1000D Series
2-32
X-Y Format
This format is useful for studying phase relationships between two signals. Channel
1 in the horizontal axis(X) and channel 2 in the vertical axis(Y), the oscilloscope uses
a none-trigger acquisition mode, data is displayed as dots.
Press horizontal MENUTime BaseX-Y to enter the X-Y mode.
Figure 2- 40
X-Y display format
NOTE: In Y-T format, all sample rates are available. But in X-Y format, 100 MSa/s
is not available. In common, deceasing the sample rate can display the wavefrom
better.
The following modes or functions will not work in X-Y format.
LA Function (DS1000D series)
Automatic Measurements
Cursor Measurements
REF and MATH Operations
Delayed Scan Mode
Vector Display Mode
Horizontal knob
Trigger Controls