RIGOL Chapter 2 Performance Verification Test
DS2000 Performance Verification Guide
d) Calculate the relative error of each scale.
10) Turn CH1 off. Repeat the above test steps to test CH2 and record the test results.
Test Record Form:
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
Vavg1
Vavg2
Calculation Result
CH1
500 μV/div
≤ 2%
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
CH2
500 μV/div
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
Note
[1]
: the calculation formula is |(Vavg1–Vavg2)–(V
out1
-V
out2
)|/Full Scale×100%; wherein, V
out1
and V
out2
are 3 and
-3 times of the current vertical scale respectively.